This model is a variable resistance unit used to apply a resistive load to generators. One unit may be used in a single phase application, or three units may be connected together to form a wye or a delta connected load and for three phase applications. The unit has sixteen toggle switch numbered clockwise from 1 to 16 and two terminals. The value of the resistances vary from 1000 W for #1 position to 18.9 W for #16 position. By placing the toggle switch on, the resistance is placed in parallel with its adjoining resistances.

OPERATION:

1. Place all the toggle switches in their ON position before connecting this unit to the generator. (All 16 switches ON means the smallest value of resistance).

2. After the generator begins to operate, increase the loud on the generator by closing each toggle switch beginning with #16 switch and proceeding in numerical order to #1 switch.